Setup Load Cell:

Parameter Current Updated
MEASURE TYPE -
CALIBRATION MODE -
MEASURE UNIT -
CELL SENSIBILITY [mV/V] -
CELL FULLSCALE [g] -
REFERENCE WEIGHT [g] -
FILTER LEVEL -
DENOISE FILTER -
RESOLUTION MODE -
STABLE DELTA WEIGHT [g] -
STABLE DELTA TIME [x100ms] -

βœ… Stable Weight Detection (Recommended):

Set STABLE DELTA WEIGHT = 2.0g and STABLE DELTA TIME = 10 (1 second) to eliminate weight jumps. Only stable readings will be sent to the monitor.

Tip: For 1g steps, also set RESOLUTION MODE = MANUAL RESOLUTION with value 1.0

πŸš€ For Maximum Responsiveness (No Filter):

Set FILTER LEVEL = 0 (2ms) and DENOISE FILTER = 0 (Disabled) for the fastest, most responsive weight readings with no filtering applied.

Live Weight Monitor (Real-Time):

OFF
SCALE WEIGHT
--- g
Unstable
Connection:
Disconnected
Updates:
0
Rate:
-- Hz
Uptime:
0s

TEST AND LOAD CELL CALIBRATION:

Measurement Value Unit
NET WEIGHT - [g]
GROSS WEIGHT - [g]
TARE WEIGHT - [g]
ADC FILTERED -
PIECES COUNT -

STATUS REGISTER

STABLE WEIGHT
FULLSCALE CELL
NEGATIVE WEIGHT

LOAD CELL CALIBRATION

Calibration Procedure:

  1. Step 1: Ensure reference weight is configured in Setup Load Cell tab
  2. Step 2: Ensure scale platform is empty and stable, then click "Tare Acquisition"
  3. Step 3: Place the reference weight on the scale and wait for stability
  4. Step 4: Once stable, click "Accept Reference Weight" to complete calibration
Connection: Disconnected
Last Update: Never